Anesthesia Machine ZAM-A16 with a flowmeter range of 0.1 to 10 L/min for each gas, offers flexibility across various clinical scenarios. It allows for customizable inspiratory-to-expiratory ratios from 4:1 to 1:8. The vaporizer integrates temperature, pressure, and flow compensation, along with a self-lock mechanism. Its separate electric and gas circuits enhance durability. Our machine offers multiple customizable triggers, including pressure and flow, for optimal ventilation support.
For adult, pediatric, and neonatal patients
Multiple ventilation modes
Rapid oxygen supply
Adjustable tidal volume and respiratory rates
Real-time graphical displays
Vaporizer with compensation and safety lock
13 alarms with visual and sound alerts
Built-in battery with 2–3 hours backup
Multilingual interface with 8 languages
Accessories Quantity Anesthetic breathing system 1pc CO₂ sensor 1pc Anesthetic vaporizer 2pcs Power cord (4m) 1pc Fuse (250V) 11pcs O₂ cell 1pc Adult anesthesia breathing circuit (reusable) 2pc Child anesthesia breathing circuit (reusable) 2pc Breathing tube 1pc Y connector (Adult) 1pc Y connector (Children) 1pc L connector 2pc Straight connector 1pc Reservoir bag 1L 1pc Reservoir bag 3L 1pc Anesthetic mask adult 1pc Anesthetic mask infant 1pc High pressure O₂ hose 1pc High pressure air hose 1pc High pressure N₂O hose 1pc Auxiliary Common Gas Outlet (ACGO) 1pc Drawers 2pc Vaporizer 2pc
Spare Cylinder yoke, 1 O₂, includes gauges
Spare Cylinder yoke, 1 N₂O, includes gauges
Anesthetic gas scavenging system
Disposable adult breathing circuit
Disposable pediatric breathing circuit
Patient monitor
Our Anesthesia Machine ZAM-A16 is used in intensive care units, operating rooms, and anesthesiology departments for adult, pediatric, and neonatal inhalation anesthesia and respiratory management.
Ventilation Mode | IPPV, A/C, PCV, SIMV, SIGH, MAUN |
---|---|
Gas Source | O₂, N₂O/O₂, AIR |
Flowmeter Range | O₂: 0.1 to 10 L/min N₂O: 0.1 to 10 L/min AIR: 0.1 to 10 L/min |
Rapid Oxygen Supply | 25 L/min to 75 L/min |
Oxygen Concentration | 15% to 100% |
Inspiratory-To- Expiratory Ratio (I:E) | 4:1 to 1:8 |
Tidal Volume (VT) | 0,20ml to 1500ml |
Minute Volume (MV) | 0 L/min to 100 L/min |
Respiratory Rate (Freq) | 1/min to 100/min |
Positive End-Expiratory Pressure (PEEP) | 0cmH₂O to 30cmH₂O |
Pressure Control (PC) | 5cmH₂O to 60cmH₂O |
Pressure Trigger (PTr) | -20 cmH₂O to 20 cmH₂O |
Flow Trigger (FTr) | OFF, 1 L/min to 30 L/min |
Inspiratory Time (Tinsp) | 0s to 12s |
Inspiratory Pause (TIP) | 0s to 6s |
Apnea Ventilation | OFF, 5s to 60s |
SIGH | 0 (off) 1/100 to 5/100 |
Gas Pressure | 280kPa to 600kPa |
Bellow Volume | 1.5 L |
Graphical Display | P-T (pressure – time) F-T (flow – time) P-V loop (pressure – volume loop) |
Alarms | AC Power Failure Alarm: Power failure or no connection Internal Battery Low Voltage: < 11.3 ±0.3V No Tidal Volume: ≤5mL within 6s High Oxygen Concentration: 19% to 100% Low Oxygen Concentration: 18% to 99% High Airway Pressure: 20cmH₂O to 100 cmH₂O Low Airway Pressure: 0 cmH₂O to 20 cmH₂O High Minute Volume: Adult (5L/min to 20L/min) Low Minute Volume: Pediatric (1L/min to 15L/min, 0 to 10L/min) Continuous Pressure: (PEEP + 1.5 kPa) over 16s Suffocation Warning: 5s to 60s no spontaneous ventilation Maximum Limited Pressure: ≤ 12.5 kPa |
Error Indicators | Fan Error: Shown on screen Oxygen Deficit: Shown on screen |
Display | 10.4"Colorful TFT LCD display screen |
Power Consumption | 85VA |
Power Supply | 100V to 240V, 50/60Hz |
Packaging Dimensions (L×W×H) | 740 × 795 × 1540 mm |
Net Weight | 52.5 kg |
Gross Weight | 119 kg |