Overview

Infant Incubator ZII-A50 is a microcomputer-controlled device with air mode and baby mode for effective incubation. It delivers air temperature control range of 25 ℃ to 37 ℃ and baby temperature control range of 34 ℃ to 37 ℃. Designed with LCD for storage, checking and modifications of the parameters like humidity and oxygen concentration level for the convenience of the user. Equipped with X ray cassette under the bassinet with low noise system and RS-232 connector for transmission of data.

Specifications :
Control mode Air mode and baby mode
Air temperature control range 25 ℃ to 37 ℃
37 ℃ to 39 ℃
Baby temperature control range 34 ℃ to 37 ℃
37 ℃ to 38 ℃
Accuracy of skin temperature sensor ± 0.2 ℃
Temperature fluctuation ≤ 0.5 ℃
Temperature uniformity ≤ 0.8 ℃
Inclination of the bassinet ± 8 °
Noise inside hood ≤ 45 dB
Humidity display range 0 % RH to 99 %RH
Humidity control range 0% RH to 90 % RH
Accuracy of humidity display ± 5% RH
Indication range of oxygen concentration 0% to 99%
Control range of oxygen concentration ± 2% (Oxygen concentration sets ≤25 %)
± 3% (Oxygen concentration sets >25 %)
Accuracy of O2 control concentration ± 4 %
Weighing range 100 to 8000 g
Resolution of weight indication 1 g
Accuracy of weight ± 1 %
Dimension (Main body) (L×W×H) 1110 × 710 × 888 mm
Dimension (Cabinet) (L×W×H) 1029 × 697 × 826 mm
Gross weight (Main body) 70 Kg
Gross weight (Cabinet) 80 Kg
Features :
  • Air mode and baby control mode for the operation of the incubator
  • LCD display for storage, checking and modification of parameters
  • Humidity and oxygen concentrations are displayed on the display for reliability
  • Temperature can be set for more than 37 ℃
  • Air circulation with outlet in front and back of the tool
  • Failure alarm indications for the high efficiency of the experiments
  • Adjustable height of the incubator for ease of the operator
  • RS-232 data interface for output of the data
  • Low noise for efficient working environment
  • Integrated sensor module and water reservoir in drawer shape
  • Adjustable inclination for the bassinet
Applications :

Infant Incubator is widely used for maintaining the desired temperature for premature babies in hospitals.