Hemodialysis Machine ZHD-A10 is an effective device that was designed to simplify dialysis and improve patient outcomes and experience. It helps to facilitate rapid and trouble-free dialysis offering a Dialysate flux of 300 to 800 mL/min. Hemodialysis is a treatment that replaces the work of the own kidneys to clear wastes and extra fluid from the blood. Intended with a special filter, dialyzer, or artificial kidney. Safety features with an in-built alarm system detect the bubble or any blood leaks ensuring accurate and reliable results.
Dialysate temperature | 34.0 °C to 39.0 °C |
Dialysate flux | 300 to 800 ml/min |
Dialysate concentration | 12.1 mS/cm to 16.0 mS/cm, ± 0.1 mS/cm |
UF rate Flow range | 0 ml/h to 4000 ml/h |
Dialysate Resolution ratio | 1 ml |
Dialysate Precision | ±30 ml/h |
Extracorporeal Venous pressure | -180 mm Hg to +600 mmHg, ±10 mm Hg |
Extracorporeal Arterial pressure | -380 mm Hg to + 400 mm Hg, ± 10 mm Hg |
TMP pressure | -180 mm Hg to + 600 mm Hg, ±20 mm Hg |
Blood pump flow range | 20 ml/min to 400 ml/min(diameter:Ф6 mm) |
Spare pump flow range | 30 ml/min to 600 ml/min(diameter:Ф8 mm) |
Resolution ratio | 1 mL |
Heparin Pump Syringe size | 20, 30, 50 mL |
Heparin pump Flow range | 0 ml/h to 10 ml/h |
Heparin pump Resolution ratio | 0.1 ml |
Water input pressure | 0.15 MPa to 0.6 MPa, 21.75 P.S.I. to 87 P.S.I |
Water input temperature | 10℃ to 30℃ |
Working environment | 10℃ to 30℃ at relative humidity of no morethan 70% |
Power | 1500 W |
Power supply | AC220V/50Hz / 60Hz, 10A |
Dimension (L×W×H) | 380 × 400 × 1380 mm |
Weight | 88 Kg |
Integrated with Intelligent Double Operation System controlling and monitoring the various parameters
Designed with a TFT LCD touch Screen with a user-friendly button interface
Offers a Dialysate temperature with a preset range of 34.0℃ to 39.0℃, ±0.5℃
Adopts Blood leak detection with Photochromic property
In-built alarm system alerts when erythrocyte specific volume is 0.32±0.02 or blood leak volume is equal to or more than 1ml per liter of dialysate
The bubble detection system detects a single air bubble volume is more than 200µl at 200ml/min blood flow with an ultrasonic alarm
Intended with conductivity composed of ±0.5% acoustic-optic
Hemodialysis machine is applicable for chronic renal failure and other blood purification treatments. This device should be used in medical units specially designed, produced, and sold for renal failure patients to receive hemodialysis, which is not allowed to be used for other purposes.
Frequently Asked Question
A typical hemodialysis session lasts 3 to 5 hours, and is usually performed three times a week. Peritoneal dialysis can be done daily, either manually or with a machine overnight.
The procedure itself is not painful, but some patients may experience discomfort from needle insertion or low blood pressure during treatment.
Most machines have backup battery systems for short-term use. In case of extended outages, patients should contact their dialysis center for assistance.
Most machines have error codes displayed on the screen. If an alarm sounds, follow the manufacturer’s instructions or contact a dialysis technician for assistance.
Your healthcare team will monitor your blood tests, fluid levels, and symptoms to ensure that dialysis is effective. Feeling better and having more energy are also good signs.